Itc VerifyNow System Platelet Reactivity Test User Manual

Browse online or download User Manual for Health and hygiene Itc VerifyNow System Platelet Reactivity Test. ITC VerifyNow System Platelet Reactivity Test User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 104
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
System
User Manual
3985 Sorrento Valley Blvd.
San Diego, California 92121 U.S.A.
Telephone (for US): 1-800-643-1640
Outside US: +1-858-643-1600
Email address: [email protected]
Website: www.accumetrics.com
Page view 0
1 2 3 4 5 6 ... 103 104

Summary of Contents

Page 1 - User Manual

SystemUser Manual3985 Sorrento Valley Blvd.San Diego, California 92121 U.S.A.Telephone (for US): 1-800-643-1640Outside US: +1-858-643-1600Email addre

Page 4

Page 97 14439.C 03/2013Index - 13VerifyNow System User Manual | USA13 IndexAabciximab 3, 5, 10, 11, 31, 32, 40, 79adenosine diphosphate 4, 83A

Page 6 - Intended Use

Page 99 14439.C 03/2013Index - 13VerifyNow System User Manual | USACustomer Support Accumetrics and its distribution partners are available to as

Page 7 - 1 Introduction

Accumetrics, Inc.3985 Sorrento Valley Blvd.San Diego, California 92121 U.S.A.Telephone (for US): 1-800-643-1640Outside US: +1-858-643-1600Email addres

Page 8 - Introduction - 1

Page 7 14439.C 03/2013Overview of System Components - 2VerifyNow System User Manual | USA2 Overview of System ComponentsThe VerifyNow System con

Page 9 - 1.2 Test Procedure

Overview of System Components - 2Page 8 14439.C 03/2013VerifyNow System User Manual | USA2.1.1 Test Device Port and CoverThe test device port hol

Page 10

Page 9 14439.C 03/2013Overview of System Components - 2VerifyNow System User Manual | USA2.1.3 Icon KeysThe four icon keys directly to the right

Page 11 - 2.1 Instrument

Overview of System Components - 2Page 10 14439.C 03/2013VerifyNow System User Manual | USA2.1.8 Printer (Accessory)The printer is an optional acc

Page 12 - Figure 2‑2 Display Screen

Page 11 14439.C 03/2013Overview of System Components - 2VerifyNow System User Manual | USA2.3 Sample Collection TubesTheVerifyNowSystemusesw

Page 13 - Figure 2‑3 Power Switch

Overview of System Components - 2Page 12 14439.C 03/2013VerifyNow System User Manual | USA2.4.2 Wet Quality Controls (WQC)Wet Quality Controls co

Page 14 - 2.2 Test Device

Page 13 14439.C 03/2013Precautions - 3VerifyNow System User Manual | USA3 PrecautionsThissectionprovidesanoverviewofsafeoperatingprocedu

Page 16

Page 15 14439.C 03/2013Reagent Storage and Handling - 4VerifyNow System User Manual | USA4 Reagent Storage and HandlingThis section describes sp

Page 17 - 3 Precautions

14439.C 03/2013VerifyNow System User Manual | USAAccumetrics Standard Terms and Conditions ... 1About thi

Page 19 - 4.1 Storage and Stability

Page 17 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USA5 Quality ControlsAccumetrics has designed the VerifyNow System wit

Page 20

Quality Controls - 5Page 18 14439.C 03/2013VerifyNow System User Manual | USAUse the following procedure to perform an EQC:NOTE: Pressing the Back

Page 21 - 5 Quality Controls

Page 19 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USA4. Remove the EQC device from the storage bay on the right side of

Page 22 - VerifyNow

Quality Controls - 5Page 20 14439.C 03/2013VerifyNow System User Manual | USA7. Iftheinstrumentisoperatingwithinitsspecications,PASS wil

Page 23 - Figure 5‑5 Remove Device

Page 21 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USAProcedure1. Locate one test device and one WQC diluent tube. 2. P

Page 24 - 5.2 Wet Quality Control

Quality Controls - 5Page 22 14439.C 03/2013VerifyNow System User Manual | USA7. Openthecover.Usingthengergrip,insertthetestdeviceinto

Page 25

Page 23 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USA10. Wait for the image of the tube to display, then insert the WQC s

Page 26 - Use bar code

Quality Controls - 5Page 24 14439.C 03/2013VerifyNow System User Manual | USA13. When the results display, record the results or print the resul

Page 27 - KEEP COVER CLOSED

Page 25 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USAcontrol material does not produce a result within the expected range

Page 28 - Figure 5‑16 Remove Device

14439.C 03/2013 VerifyNow System User Manual | USA7 Retrieving Test Results and Performance Data ... 437.1 Retrieving Patie

Page 29

Quality Controls - 5Page 26 14439.C 03/2013VerifyNow System User Manual | USAQC26-Feb-201214:57:00 104AVerifyNowSystem®26-Feb-201214:57:00Figure 5

Page 30

Page 27 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USAFigure 5‑22 Scan Bar CodeBar CodeWQC P2Y12Use bar codescanner to rea

Page 31

Quality Controls - 5Page 28 14439.C 03/2013VerifyNow System User Manual | USA12. Replace the cap on the diluent tube by pressing and turning simul

Page 32 - Figure 5‑27 Close Cover

Page 29 14439.C 03/2013Quality Controls - 5VerifyNow System User Manual | USAFigure 5‑28 Test ProgressWQC P2Y1226-Feb-201214:57:00KEEP COVER CLOS

Page 33

Quality Controls - 5Page 30 14439.C 03/2013VerifyNow System User Manual | USA18. At the conclusion of the test, the resulting value will be displa

Page 34

Page 31 14439.C 03/2013Patient Testing - 6VerifyNow System User Manual | USA6 Patient TestingThe VerifyNow System has been developed to provide

Page 35 - 6 Patient Testing

Patient Testing - 6Page 32 14439.C 03/2013VerifyNow System User Manual | USA6.2 Sample CollectionSamples can be collected from an indwelling cath

Page 36 - 6.2 Sample Collection

Page 33 14439.C 03/2013Patient Testing - 6VerifyNow System User Manual | USA6. Gently invert the sample tube 5 times to ensure complete mixing o

Page 37

Patient Testing - 6Page 34 14439.C 03/2013VerifyNow System User Manual | USA6.3 Sample HandlingFreshwholebloodsamplesintheappropriateparti

Page 38

Page 35 14439.C 03/2013Patient Testing - 6VerifyNow System User Manual | USA3. From the main menu, enter Operator ID and Password, if requir

Page 40 - Patient Testing - 6

Patient Testing - 6Page 36 14439.C 03/2013VerifyNow System User Manual | USANOTE: If you make a mistake while entering a text character, press the

Page 41

Page 37 14439.C 03/2013Patient Testing - 6VerifyNow System User Manual | USA9. If the test deviceistherstfromanewlot,theinstrumentwil

Page 42

Patient Testing - 6Page 38 14439.C 03/2013VerifyNow System User Manual | USA12. Close the cover to the test port (Figure 6-11).Figure 6‑11 Close C

Page 43 - Page 39 14439.C 03/2013

Page 39 14439.C 03/2013Patient Testing - 6VerifyNow System User Manual | USAFigure 6‑13 Remove Device Press the Next key to return to the main

Page 44 - Figure 6‑16 IIb/IIIa Results

Patient Testing - 6Page 40 14439.C 03/2013VerifyNow System User Manual | USA%26-Feb-201214:57:00VerifyNow® P2Y12PRU:164Figure 6‑15 P2Y12 ResultsTh

Page 45 - INHIBITION

Page 41 14439.C 03/2013Patient Testing - 6VerifyNow System User Manual | USAUse the following procedure to calculate percent inhibition.NOTE: A s

Page 46

Patient Testing - 6Page 42 14439.C 03/2013VerifyNow System User Manual | USA6.6 Shutting down the InstrumentPrior to shutting down the instrument

Page 47

Page 43 14439.C 03/2013Retrieving Test Results - 7VerifyNow System User Manual | USA7 Retrieving Test Results and Performance DataThe instrument

Page 48 - 14:57:00

Retrieving Test Results - 7Page 44 14439.C 03/2013VerifyNow System User Manual | USA3. Whenthefollowingscreendisplays(Figure7-1),enterthe

Page 49 - Assay Results

Page 45 14439.C 03/2013Retrieving Test Results - 7VerifyNow System User Manual | USA7.1.3 Test Results LogThe screen displays a scrollable log o

Page 50 - Figure 7‑4 EQC Results Log

Page 1 14439.C 03/2013VerifyNow System User Manual | USAAccumetrics Standard Terms and ConditionsLimited Warranty and Disclaimer. Accumetrics war

Page 51 - Retrieving Test Results - 7

Retrieving Test Results - 7Page 46 14439.C 03/2013VerifyNow System User Manual | USAIn addition to viewing earlier results, icons keys enable resu

Page 52 - Figure 7‑5 WQC Results Log

Page 47 14439.C 03/2013Retrieving Test Results - 7VerifyNow System User Manual | USAFeature DescriptionEQC ResultDisplays the result as either a

Page 53

Retrieving Test Results - 7Page 48 14439.C 03/2013VerifyNow System User Manual | USA7.2.2 Wet Quality Control Results LogThe cursor is positioned

Page 54 - 7.4 View Event Log

Page 49 14439.C 03/2013Retrieving Test Results - 7VerifyNow System User Manual | USA3. Press the Next key four times. Refer to the Appendix fo

Page 55

Retrieving Test Results - 7Page 50 14439.C 03/2013VerifyNow System User Manual | USA7.4 View Event LogThe instrument stores the last 50 error/att

Page 56

Page 51 14439.C 03/2013Retrieving Test Results - 7VerifyNow System User Manual | USAIcon DescriptionPrint ResultScroll to the selected record, us

Page 58

Page 53 14439.C 03/2013Instrument Settings - 8VerifyNow System User Manual | USA8 Instrument SettingsThissectiondescribestheinstrumentcong

Page 59 - 8.4 Set Time Out

Instrument Settings - 8Page 54 14439.C 03/2013VerifyNow System User Manual | USA5. Press the Lighten key or the Darken key to adjust the scre

Page 60 - Figure 8‑4 Time Out Setting

Page 55 14439.C 03/2013Instrument Settings - 8VerifyNow System User Manual | USA8.3 Set the Date and TimeThedateandtimeisinitiallycongure

Page 61 - 8.5 Enable Patient ID

Page 2 14439.C 03/2013VerifyNow System User Manual | USAAbout this ManualThis user manual is the place to begin if you are not familiar with the V

Page 62

Instrument Settings - 8Page 56 14439.C 03/2013VerifyNow System User Manual | USAuser login, if applicable. This time period can vary based on the

Page 63 - 8.7 Set EQC Frequency

Page 57 14439.C 03/2013Instrument Settings - 8VerifyNow System User Manual | USA8.5 Enable Patient IDThe VerifyNow Instrument supports patient r

Page 64 - 8.8 Enable Printer

Instrument Settings - 8Page 58 14439.C 03/2013VerifyNow System User Manual | USA1. From the main menu, enter Operator ID and Password, if req

Page 65 - Figure 8‑8 Printer Setting

Page 59 14439.C 03/2013Instrument Settings - 8VerifyNow System User Manual | USA26-Feb-201214:57:00Use Password: YesOper Pswd Auth444 1323 M

Page 66

Instrument Settings - 8Page 60 14439.C 03/2013VerifyNow System User Manual | USANOTE: When the established interval has elapsed, the user is locke

Page 67 - 9 Cleaning and Maintenance

Page 61 14439.C 03/2013Instrument Settings - 8VerifyNow System User Manual | USA26-Feb-201214:57:00Version: X00191Use PtId> YesPtId Lth:3S/N:

Page 69 - 9.3 Replace the Fan Filter

Page 63 14439.C 03/2013Cleaning and Maintenance - 9VerifyNow System User Manual | USA9 Cleaning and MaintenanceThe VerifyNow Instrument does not

Page 70 - Figure 9‑4 Remove Fan Filter

Cleaning and Maintenance - 9Page 64 14439.C 03/2013VerifyNow System User Manual | USA3. Open the cover and insert the cleaning device into the te

Page 71 - ERROR 20

Page 65 14439.C 03/2013Cleaning and Maintenance - 9VerifyNow System User Manual | USAProcedureUse a recommended cleaning solution on the exterior

Page 72 - 10.1 Messages

Page 3 14439.C 03/2013Introduction - 1VerifyNow System User Manual | USA1 IntroductionBefore you begin to use the VerifyNow System, it is approp

Page 73 - Section 8.3)

Cleaning and Maintenance - 9Page 66 14439.C 03/2013VerifyNow System User Manual | USANOTE: There are no screws to be removed.Figure 9‑4 Remove Fan

Page 74

Page 67 14439.C 03/2013Troubleshooting - 10VerifyNow System User Manual | USA10 TroubleshootingOccasionally, an unanticipated event occurs which

Page 75

Troubleshooting - 10Page 68 14439.C 03/2013VerifyNow System User Manual | USAElectronics Diagnostics- During an EQC, the system monitors the elect

Page 76

Page 69 14439.C 03/2013Troubleshooting - 10VerifyNow System User Manual | USAIssue Description ResolutionMessage – Non-matching baseline values (

Page 77 - 10.3 Alarm Messages

Troubleshooting - 10Page 70 14439.C 03/2013VerifyNow System User Manual | USANo. Issue Description Resolution4Error – Test device spot code The in

Page 78 - 10.4 EQC Diagnostic Messages

Page 71 14439.C 03/2013Troubleshooting - 10VerifyNow System User Manual | USANo. Issue Description Resolution12Attention - Tube or test device re

Page 79 - Troubleshooting - 10

Troubleshooting - 10Page 72 14439.C 03/2013VerifyNow System User Manual | USANo. Issue Description Resolution21Error - Low proximal voltage Test o

Page 80

Page 73 14439.C 03/2013Troubleshooting - 10VerifyNow System User Manual | USANo. Issue Description Resolution28Attention - Signal saturation High

Page 81 - 11 Appendix

Troubleshooting - 10Page 74 14439.C 03/2013VerifyNow System User Manual | USANo. Issue Description Resolution104Alarm 104Heater Timeout During Hea

Page 82 - 11.4 Glossary

Page 75 14439.C 03/2013Troubleshooting - 10VerifyNow System User Manual | USAParameter Issue Description ResolutionV2 Maximum leak rate This even

Page 83 - Appendix - 11

Introduction - 1Page 4 14439.C 03/2013VerifyNow System User Manual | USA1.1.1 Aspirin Test Acetylsalicylic acid (aspirin)hasasignicantantipl

Page 85 - Page 81 14439.C 03/2013

Page 77 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USA11 Appendix11.1 Technical SpecicationsFeature SpecicationDimensions (Wx

Page 86

Appendix - 11Page 78 14439.C 03/2013VerifyNow System User Manual | USA11.3 System ComponentsManufacturer Catalog No.Description86000VerifyNow Sys

Page 87 - 11.5 Software Icons

Page 79 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USATerm DescriptionAdenosine diphosphateADP. When secreted by activated platel

Page 88

Appendix - 11Page 80 14439.C 03/2013VerifyNow System User Manual | USATerm DescriptionDetection wellThe detection wells of an test device contain

Page 89 - Page 85 14439.C 03/2013

Page 81 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USATerm DescriptionInhibitor, glycoprotein (GP) IIb/IIIaGP IIb/IIIa inhibitors

Page 90 - Run a wet quality control

Appendix - 11Page 82 14439.C 03/2013VerifyNow System User Manual | USATerm DescriptionPAR-4 receptorPAR-4 is one of four protease-activated recept

Page 91

Page 83 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USATerm DescriptionPouch The white test device package.Power indicatorWhen pow

Page 92 - Instrument Settings and Logs

Appendix - 11Page 84 14439.C 03/2013VerifyNow System User Manual | USAIcon Name DescriptionAdd/Update Record Select this icon to add an operator’s

Page 93

Page 85 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USAIcon Name DescriptionPrint All Select this icon to print the records starti

Page 94 - Month_________ Year______

Page 5 14439.C 03/2013Introduction - 1VerifyNow System User Manual | USA1.1.3 IIb/IIIa Test Thenalcommonpathwaytoplateletaggregationinvo

Page 95 - Page 91 14439.C 03/2013

Appendix - 11Page 86 14439.C 03/2013VerifyNow System User Manual | USA11.6 Instrument Software – Test and QCRetry scanMain menuStart screen Inse

Page 96

Page 87 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USARetry scanMain menuStart screen Insert as saydeviceScan barcode Insert tub

Page 97 - Page 93 14439.C 03/2013

Appendix - 11Page 88 14439.C 03/2013VerifyNow System User Manual | USA11.7 Main screenMaintenanceMenu 1MaintenanceMenu 2MaintenanceMenu 3Maintenan

Page 98

Page 89 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USAMain screenMaintenanceMenu 1MaintenanceMenu 2MaintenanceMenu 3MaintenanceMe

Page 99 - 12 References

Appendix - 11Page 90 14439.C 03/2013VerifyNow System User Manual | USA11.8 VerifyNow Instrument Cleaning and Maintenance Schedule Month_________

Page 100

Page 91 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USA11.9 Compliance with CLIA Control RequirementsThe CLIA Regulation, section

Page 101 - 13 Index

Appendix - 11Page 92 14439.C 03/2013VerifyNow System User Manual | USAprocedures or changing instrument parameters).• As an additional security f

Page 102

Page 93 14439.C 03/2013Appendix - 11VerifyNow System User Manual | USAThe test device internal controls can detect failures of the reagent system

Page 103 - Customer Support

Appendix - 11Page 94 14439.C 03/2013VerifyNow System User Manual | USA• EachinstrumentistestedwithapreviouslyqualiedlotofVerifyNowtes

Page 104

Page 95 14439.C 03/2013References - 12VerifyNow System User Manual | USA12 References1. Centers for Disease Control. Perspectives in Disease Pr

Comments to this Manuals