RxDx
®
Analysis Module
Heparin and Protamine Dosing
5WRRNGOGPVVQ*'/1%*410Response
9JQNG$NQQF%QCIWNCVKQP5[UVGO
Operator’s Manual
English
TABLE OF CONTENTS
+06'0&'&75' 
57//#4;#0&':2.#0#6+10
24+0%+2.'51(12'4#6+10 
#66'06+10.#$'. 
)'66+0)56#46'& 
5'66+0)572'48+514126+105
12'4#6+10
37#.+6;%10641. 
4'57.65/#0#)'/'06 
4'21465  
4'('4'0%'5  
+0&': 
This manual is published by International Technidyne Corporation (ITC) for use with the HEMOCHRON
®
Response ver. 2.00 or higher.
Questions or comments regarding the content of this manual can be directed to the address at the back of
this manual or to your ITC representative.
HEMOCHRON
®
and RxDx
®
are registered trademarks of ITC.
Celite
®
is a registered trademark of Celite Corporation.
©2003, 2004. This document is the copyright of ITC and must not be copied or reproduced in any form
without prior consent. ITC reserves the right to make technical improvements to this equipment and
documentation without prior notice as part of a continuous program of product development.
Page view 0
1 2 3 4 5 6 ... 23 24

Summary of Contents

Page 1

RxDx® Analysis Module Heparin and Protamine Dosing 5WRRNGOGPVVQ*'/1%*410Response9JQNG$NQQF%QCIWNCVKQP5[UVGOOperator’s Manual Englis

Page 2

10 Selecting a Different Case or Modifying the PID for the Selected Case 1. Display the Patient Info screen: 2. Press 1 to display the screen for en

Page 3

11 6. Use the numeral keys to enter the patient weight (from 20 to 300 kilograms, or from 44.1 to 666.3 pounds, depending upon the units chosen). If

Page 4

12 Determining the Bolus Heparin Dose Once the baseline ACT result for a patient is known, the bolus heparin dose can be determined, either by perfor

Page 5

13 5. Press YES. The information is stored and the Additional Heparin Dose screen is displayed: Note: AddlHep is the calculated additional heparin do

Page 6

14 6. Press YES. The protamine dose information is stored and the RxDx screen is displayed. If YES is not pressed, the dose will not be stored. Dete

Page 7

15 Determining Circulating Heparin Level (in mg/kg) Using a HiTT Test Tube When Not in the RxDx Mode Note: Refer to the appropriate test package inser

Page 8

16 Appending User Notes User Notes can be tagged to a test result to denote additional information. When a note is appended, the numeral corresponding

Page 9

17 QUALITY CONTROL Routine quality control testing should be part of a comprehensive quality assurance program. Quality control testing of the HEMOCHR

Page 10

18 RESULTS MANAGEMENT Up to ten (10) RxDx cases can be contained on a HEMOCHRON Response with an activated HEMOCHRON RxDx Analysis Module. Note: If

Page 11

19 Printing the Test Record, Patient Record, Case List, or Bolus Dose Report 1. Return to the main RxDx menu, and press the PRINT key to display the R

Page 12

2 INTENDED USE The HEMOCHRON® RxDx® Analysis Module is a supplementary module for the HEMOCHRON Response Whole Blood Coagulation System that provides

Page 13

20 REPORTS Data from an RxDx case that was run on the HEMOCHRON Response can be retrieved and reported at any time. Seven RxDx reports are available:

Page 14

21 Displaying and Printing a Report Display and/or print various reports as summarized below: Report Description Patient Record Display the Patient

Page 16

23 INDEX Additional Heparin Dose calculation...4 determination ...

Page 17

*410

Page 18

3 PRINCIPLES OF OPERATION Refer to the HEMOCHRON Response Whole Blood Coagulation System Operator's Manual for a complete description of the HEMO

Page 19

4 Blood Volume Calculation Blood volume can either be directly entered, or it can be calculated from the patient's height and weight. The system

Page 20

5 Calculation of Heparin Level The heparin level will be calculated automatically whenever a PRT, PDA-O or HiTT test is performed. For the PRT and P

Page 21

6 ATTENTION LABEL An attention label on the rear of the HEMOCHRON Response instrument alerts users to accompanying documentation: Before using the HE

Page 22

7 SETTING SUPERVISOR OPTIONS The supervisor can specify default values for heparin units, target ACT, heparin and protamine dose adjustments, and pat

Page 23

8 Specifying the Default Percentage for Adjusting the Dose The percentage dose adjustment is the factor by which the calculated dose is increased pri

Page 24 - *410

9 OPERATION Entering RxDx The RxDx mode of operation is entered by any of the following: • Selecting RxDx from the ID Selections menu. • Inserting a